Microbial enumeration
2021
It meets the requirements of the tests for absence of
Escherichia coli in 10 g. The total aerobic microbial count does not exceed 1000 cfu per g or mL, and the total combined molds and yeasts count does not exceed 100 cfu per g or mL.
Chromatographic purity and limit of dimethyl sulfoxide
Standard stock solution
Dissolve an accurately weighed quantity of
USP Dimethyl Sulfoxide RS in methanol, and dilute quantitatively, and stepwise if necessary, with methanol to obtain a solution having a known concentration of about 1.0 mg of
USP Dimethyl Sulfoxide RS per mL.
Sensitivity check solution
Dilute the Standard stock solution with methanol to obtain a solution having a concentration of 2.0 µg per mL.
Test solution
Transfer about 100 mg of methylsulfonylmethane, accurately weighed, to a 50-mL volumetric flask, dissolve in and dilute with methanol to volume, and mix. Sonicate at 50

for 1 minute, allow to cool to room temperature, and mix.
Chromatographic system (see Chromatography
621
)
The gas chromatograph is equipped with a flame-ionization detector and a 0.53-mm × 30-m capillary column coated with a 5-µm phase G2. The column temperature is maintained at 120

. Helium is used as the carrier gas, flowing at a rate of 5.0 mL per minute. The split ratio is 2:1. The injection port temperature is maintained at 250

, and the detector temperature is maintained at 250

. Chromatograph the
Sensitivity check solution: the dimethyl sulfoxide peak must be greater than 10 times the noise height, the noise height being determined by a suitable procedure. Chromatograph the
System suitability solution, and record the peak responses as directed for
Procedure: the resolution,
R, between dimethyl sulfoxide and methylsulfonylmethane is not less than 2.0.
Procedure
Inject a volume (about 1 µL) of the
Test solution into the chromatograph, record the chromatogram, and measure the responses for the major peaks. Calculate the percentage of each impurity in the portion of Methylsulfonylmethane taken by the formula:
100(ri / rs)
in which
ri is the response of each individual impurity in the chromatogram of the
Test solution; and
rs is the sum of the responses of all of the peaks other than the solvent peak: not more than 0.1% of dimethyl sulfoxide is found; not more than 0.05% of any other individual impurity is found; and the sum of all impurities, including dimethyl sulfoxide, is not more than 0.2%.
Assay
Diluent
Transfer about 950 mL of methanol to a 1-L volumetric flask. Add 0.60 mL of di(ethylene glycol) methyl ether, and dilute with methanol to volume.
Standard preparation
Dissolve an accurately weighed quantity of
USP Methylsulfonylmethane RS in
Diluent, and dilute quantitatively, and stepwise if necessary, with
Diluent to obtain a solution having a known concentration of about 0.4 mg per mL. Sonicate at 50

for 1 minute, allow to cool to room temperature, and mix.
Assay preparation
Transfer about 40 mg of Methylsulfonylmethane, accurately weighed, to a 100-mL volumetric flask, dissolve in and dilute with
Diluent to volume, and mix. Sonicate at 50

for 1 minute, allow to cool to room temperature, and mix.
Chromatographic system (see Chromatography
621
)
The gas chromatograph is equipped with a flame-ionization detector and a 0.53-mm × 30-m capillary column coated with a 5-µm phase G2. The column temperature is maintained at 120

. Helium is used as the carrier gas, flowing at a rate of 5.0 mL per minute. The split ratio is 2:1. The injection port and detector temperatures are maintained at 250

. Chromatograph the
Standard preparation, and record the peak responses as directed for
Procedure: the relative standard deviation for replicate injections is not more than 2.0%.
Procedure
Separately inject equal volumes (about 1 µL) of the
Standard preparation and the
Assay preparation into the chromatograph, record the chromatograms, and measure the responses for the major peaks. Calculate the quantity, in mg, of C
2H
6O
2S in the portion of Methylsulfonylmethane taken by the formula:
100C(RU / RS)
in which
C is the concentration, in mg per mL, of
USP Methylsulfonylmethane RS in the
Standard preparation; and
RU and
RS are the peak response ratios of methylsulfonylmethane to di(ethylene glycol) methyl ether obtained from the
Assay preparation and the
Standard preparation, respectively.