 1230
1230 WATER FOR HEALTH APPLICATIONS
 WATER FOR HEALTH APPLICATIONS
| Element or Compound | Maximum Concentration (mg/L) | 
| Calcium | 2 (0.1 mEq/L) | 
| Magnesium | 4 (0.3 mEq/L) | 
| Potassium | 8 (0.2 mEq/L) | 
| Sodium | 70 (3.0 mEq/L) | 
| Antimony | 0.006 | 
| Arsenic | 0.005 | 
| Barium | 0.10 | 
| Beryllium | 0.0004 | 
| Cadmium | 0.001 | 
| Chromium | 0.014 | 
| Lead | 0.005 | 
| Mercury | 0.0002 | 
| Selenium | 0.09 | 
| Silver | 0.005 | 
| Aluminum | 0.01 | 
| Chloramines | 0.10 | 
| Free chlorine | 0.50 | 
| Copper | 0.10 | 
| Fluoride | 0.20 | 
| Nitrate (as N) | 2.00 | 
| Sulfate | 100.00 | 
| Thallium | 0.002 | 
| Zinc | 0.10 | 
| 
* 
Reprinted with permission from ANSI/AAMI RD62: 2001, Water treatment equipment for hemodialysis applications, copyright Association for the Advancement of Medical Instrumentation, Arlington, VA.
 | |
 and 35
 and 35 , for no less than 48 hours. Sampling the water should be done at the end of the water purification cascade at the point where the water enters the dialysis equipment. Samples should be assayed within 30 minutes of collection or immediately refrigerated and then assayed within 24 hours of collection. Quantification of bacterial endotoxins is performed using the Limulus Amebocyte Lysate (LAL) clotting method or any other LAL test found in the USP general test chapter Bacterial Endotoxins Test
, for no less than 48 hours. Sampling the water should be done at the end of the water purification cascade at the point where the water enters the dialysis equipment. Samples should be assayed within 30 minutes of collection or immediately refrigerated and then assayed within 24 hours of collection. Quantification of bacterial endotoxins is performed using the Limulus Amebocyte Lysate (LAL) clotting method or any other LAL test found in the USP general test chapter Bacterial Endotoxins Test  85
85 .
. 1231
1231 ).
).